| Model Number |
MNC30013 |
| Medical Device Problem Codes |
Retraction Problem (1536); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Review of device history records including raw material inspection, in process & finished product inspection does not reveal any discrepancy relevant to the batch under investigation, which confirms that there is no indication of a product related quality deficiency associated to this batch.Additionally, a review of the complaint history identified no other incidents reported from this lot and the control sample test results confirm that there was no indication of a product quality deficiency.Based on the available information and without concomitant device, it is not possible to draw a clinical conclusion between the device and the reported event.Neither the dhr review nor the complaint history including control sample analysis results suggest that the reported failures could be related to design or manufacturing process.Please note: the concomitant device is available, but not yet returned to the manufacturer.Device evaluation anticipated, but not yet begun.
|
| |
|
Event or Problem Description
|
|
It was reported that a 3.0x13 mozee nc was difficult to remove from sion blue wire.No patient harm or injury was reported.The device was stored and handled as per the ifu.There was no damage noted to the packaging of the device and the device was removed from the package without encountering any difficulty.There was no difficulty in removing the product from the hoop and in removing the protective balloon cover.The product was normally prepped following the ifu without using any excess force.The product was inspected prior to use and appeared to be normal.No anomalies were noted during or after the device was prepped.No visible signs of kinks or damages were noted prior to inserting the product into the patient.No resistance i friction was experienced while inserting the balloon through the rotating hemostatic valve and while crossing the target lesion.No visible signs of kink or damage were noted anytime during the procedure and no unusual force was applied anytime during the procedure.There was no difficulty encountered when connecting the hub to the indeflator and the device was not advanced with the indeflator connected to the hub.The balloon was inflated normally and complete deflation of the balloon was observed.Also the catheter was never in an acute bend.Additionally, no kink or other damages were noted after the device was removed from the patient and the product looked normal.
|
| |
|
Additional Manufacturer Narrative
|
|
Review of device history records including raw material inspection, in process & finished product inspection does not reveal any discrepancy relevant to the batch under investigation, which confirms that there is no indication of a product related quality deficiency associated to this batch.Additionally, a review of the complaint history identified no other incidents reported from this lot and the control sample test results also confirm that there was no indication of a product quality deficiency.Based on the available information there is no evidence to suggest that the event was design or manufacturing related.Neither the dhr review nor the returned product analysis suggests that the reported failures could be related to the manufacturing process.- attachment: [complaint evaluation report_mnc-pc18-026.Pdf, complaint evaluation report_mnc-pc18-026_returned product analysis.Pdf].
|
| |
|
Event or Problem Description
|
|
It was reported that a 3.0x13 mozee nc was difficult to remove from sion blue wire.No patient harm or injury was reported.The device was stored and handled as per the ifu.There was no damage noted to the packaging of the device and the device was removed from the package without encountering any difficulty.There was no difficulty in removing the product from the hoop and in removing the protective balloon cover.The product was normally prepped following the ifu without using any excess force.The product was inspected prior to use and appeared to be normal.No anomalies were noted during or after the device was prepped.No visible signs of kinks or damages were noted prior to inserting the product into the patient.No resistance i friction was experienced while inserting the balloon through the rotating hemostatic valve and while crossing the target lesion.No visible signs of kink or damage were noted anytime during the procedure and no unusual force was applied anytime during the procedure.There was no difficulty encountered when connecting the hub to the indeflator and the device was not advanced with the indeflator connected to the hub.The balloon was inflated normally and complete deflation of the balloon was observed.Also the catheter was never in an acute bend.Additionally, no kink or other damages were noted after the device was removed from the patient and the product looked normal.
|
| |
|