• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD PERIPHERAL VASCULAR, INC. ATLAS PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD PERIPHERAL VASCULAR, INC. ATLAS PTA DILATATION CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number AT75186
Medical Device Problem Codes Retraction Problem (1536); Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No hospital/medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device has been returned to the manufacturer for evaluation.The investigation of the reported event is currently underway.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a calcified lesion in the left common iliac vein, the pta balloon allegedly ruptured at 10 atm during the second inflation attempt.It was further reported that there was difficulty retracting the balloon through the sheath.The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records were reviewed with special attention to the raw materials, subassemblies, manufacturing process and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was returned for evaluation.An attempt was made to inflate the device and a longitudinal rupture was noted in the balloon.Therefore, the investigation is confirmed for a longitudinal rupture.However, the investigation is inconclusive for the reported retraction issue as the device was unable to be fully functionally tested due to the rupture, and as the user's introducer sheath was not returned with the device.Per the reported event details, the target lesion was calcified and a wall stent was present during the procedure.Therefore, it is possible that patient and/or procedural issues contributed to the event.However, the definitive root cause for the reported or identified issues could not be determined based upon the available information.It is also unknown if the original sheath used in the procedure contributed to the reported retraction issue.Labeling review: the review of the ifu (instructions for use), indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H11: h3, h6 (results 1, conclusion 1) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during an angioplasty procedure of a calcified lesion in the left common iliac vein, the pta balloon allegedly ruptured at 10atm during the second inflation attempt.It was further reported that there was difficulty retracting the balloon through the sheath.The pta balloon was exchanged over the guidewire for another that was used to complete the procedure.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS PTA DILATATION CATHETER
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key8151245
Report Number2020394-2018-02184
Device Sequence Number13859400
Product Code LIT
UDI-Device Identifier00801741062506
UDI-Public(01)00801741062506
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K120971
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/28/2021
Device Model NumberAT75186
Device Catalogue NumberAT75186
Device Lot NumberGFCV2730
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/27/2018
Initial Date Received by Manufacturer 11/14/2018
Supplement Date Received by Manufacturer01/24/2019
Initial Report FDA Received Date12/11/2018
Supplement Report FDA Received Date01/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age61 YR
Patient Weight68
-
-