• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. SUPER MULTIVAC 50 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. SUPER MULTIVAC 50 IFS; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Catalog Number ASH4830-01
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/04/2018
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during knee scope, the face/electrode of the wand came off in the joint.All pieces were removed.There was not delay in the case and back up device was available.Not patient injury was reported.
 
Additional Manufacturer Narrative
Adverse event or product problem, outcomes attributed to adverse event: information added.Type of reportable event corrected.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.There was a relationship found between the returned device and the reported incident.Visual inspection under magnification shows the electrodes have been detached with jagged edges on the remaining electrode legs.There is also discoloration on the adhesive and scuff marks on the spacer.There is a significant amount of scratch marks on the exposed shaft and black shrink tube near the tip of the wand.The cable has been severed prior to being received for evaluation; therefore, a functional test could not be conducted.There are no manufacturing abnormalities visually observed with the returned device.The complaint has been visually verified and the root cause was more than likely associated with the device potentially coming into contact with a metal object such as a cannula.Other factors that could have led to tip detachment includes: mechanical displacement of tissue through applied force or using the device as a lever to enlarge a surgical site or gain access to tissue.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation.A relationship, if any, between the subject device and the reported event could not be determined since the product was not returned for evaluation.Visual inspection and functional testing could not be performed because the device in question was not returned for evaluation.Thus, the complaint could not be verified, nor could a root cause be determined with confidence.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.There were no indications during manufacturing record review that would suggest that the device did not meet product specifications upon release into distribution.
 
Event or Problem Description
It was reported that during knee arthroscopy/meniscectomy, the face/electrode of the wand came off in the joint.X-rays were performed confirming that all parts were found and retrieved.A final lavage was performed looking at all compartments and the gutters.There was a significant delay in the case and back up device was available.Not patient injury was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPER MULTIVAC 50 IFS
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key8151375
Report Number3006524618-2018-00638
Device Sequence Number2010004
Product Code GEI
UDI-Device Identifier00817470004908
UDI-Public00817470004908
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K083306
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 03/28/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/17/2020
Device Catalogue NumberASH4830-01
Device Lot Number2008296
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2019
Initial Date Received by Manufacturer 12/04/2018
Supplement Date Received by Manufacturer01/22/2019
01/22/2019
01/23/2019
03/28/2019
Initial Report FDA Received Date12/11/2018
Supplement Report FDA Received Date01/23/2019
01/23/2019
02/06/2019
03/28/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age46 YR
Patient Weight89
-
-