Model Number 44150
Device Problems Retraction Problem (1536); Material Split, Cut or Torn (4008)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/15/2018
Event Type  malfunction  
Event Description
It was reported that the sheath became torn.A 190cm filterwire ez was selected for use.During preparation, the device was taken out from the hoop and flushed.When the device filter was retracted in the sheath, it was noted that there was resistance felt and it could not be retracted.Furthermore, the wire became bent when it was forcibly retracted and the sheath was torn.The procedure was completed with another of the same device.No patient complications were reported and the patient's condition was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FILTERWIRE EZ
Type of Device
TEMPORARY CAROTID CATHETER FOR EMBOLIC CAPTURE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8151525
MDR Text Key130048563
Report Number2134265-2018-63817
Device Sequence Number1
Product Code NTE
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063313
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/30/2020
Device Model Number44150
Device Catalogue Number44150
Device Lot Number0022450513
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/19/2018
Initial Date FDA Received12/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-