| Model Number |
71702-01 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Hypoglycemia (1912)
|
| Date of Event |
11/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
A customer reported that her adc freestyle libre sensor fell off prematurely, after 5 days of wear.On (b)(6) 2018 the customer felt light headed and weak.The customer was seen by her doctor who gave her "some juice" as treatment to raise her blood glucose level.There was no report of death of permanent injury associated with this event.
|
| |
|
Event or Problem Description
|
|
A customer reported that her adc freestyle libre sensor fell off prematurely, after 5 days of wear.On (b)(6) 2018 the customer felt light headed and weak.The customer was seen by her doctor who gave her "some juice" as treatment to raise her blood glucose level.There was no report of death of permanent injury associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.A dhr (device history review) for the libre sensor kit was reviewed and the dhr showed the libre sensor kit passed all tests prior to release.All review activities conducted above, including but not limited to the final release testing specifically associated with the manufacture of this product, are sufficient information in order to show if the product has met specifications prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.
|
| |
|
Search Alerts/Recalls
|