Medications - aspirin, amlodipine, omeprazole, repatha.The review of the manufacturing paperwork verified that this lot met all pre-release specifications.According to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and/or require intervention include, but are not limited to component migration, endoleak and aneurysm enlargement.
|
|
On (b)(6) 2014, the patient underwent endovascular treatment of an abdominal aortic aneurysm with gore® excluder® aaa endoprostheses.On an unknown date, a follow-up computed tomography scan identified a suspected proximal type i endoleak and aneurysm enlargement of an unknown amount.The cause of the proximal type i endoleak is reportedly due to distal migration (amount unknown) of the trunk-ipsilateral leg component.The migration was reported caused by dilatation of the aortic neck causing a loss of seal.On (b)(6) 2018, a reintervention was performed whereby an aortic extender component was implanted for proximal extension off of the trunk-ipsilateral leg component.It was reported the endoleak was resolved and the patient tolerated the procedure.
|