• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3); SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Catalog Number RMT281414
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 12/07/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medications - aspirin, amlodipine, omeprazole, repatha.The review of the manufacturing paperwork verified that this lot met all pre-release specifications.According to the gore® excluder® aaa endoprosthesis instructions for use, adverse events that may occur and/or require intervention include, but are not limited to component migration, endoleak and aneurysm enlargement.
 
Event or Problem Description
On (b)(6) 2014, the patient underwent endovascular treatment of an abdominal aortic aneurysm with gore® excluder® aaa endoprostheses.On an unknown date, a follow-up computed tomography scan identified a suspected proximal type i endoleak and aneurysm enlargement of an unknown amount.The cause of the proximal type i endoleak is reportedly due to distal migration (amount unknown) of the trunk-ipsilateral leg component.The migration was reported caused by dilatation of the aortic neck causing a loss of seal.On (b)(6) 2018, a reintervention was performed whereby an aortic extender component was implanted for proximal extension off of the trunk-ipsilateral leg component.It was reported the endoleak was resolved and the patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (C3)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key8151632
Report Number2017233-2018-00754
Device Sequence Number2125538
Product Code MIH
UDI-Device Identifier00733132610716
UDI-Public00733132610716
Combination Product (Y/N)N
Initial Reporter StateME
Initial Reporter CountryUS
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2016
Device Catalogue NumberRMT281414
Device Lot Number12126842
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age68 YR
Patient Weight88
-
-