• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HEMODIALYSIS SYSTEM W/BIBAG; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HEMODIALYSIS SYSTEM W/BIBAG; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 190766
Device Problem Thermal Decomposition of Device (1071)
Patient Problem No Patient Involvement (2645)
Event Date 12/07/2018
Event Type  malfunction  
Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a burned power cord.The burned power cord was noticed during daily setup.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The cord was the original fresenius part on the machine.The biomed reported that there was no melting, burning smell, smoke, spark, flame, or arcing observed.The biomed reported that the machine has not had any past problems with failing the electrical leakage test and that there was no damage observed on any other components.The machine has approximately 12,000 hours and is plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet.The biomed replaced the plug, which resolved the issue.The biomed reported that the machine has been returned to service without issue.The power plug was reported have been discarded and is not available for manufacturer evaluation.
 
Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.During an onsite repair, the user facility¿s biomedical technician resolved the reported issue by replacing the burnt power cord.The biomedical technician reported that the machine had been returned to service without issue.A serial number search in the complaint handling system found no complaint issues with the same symptom within 90 days of the notified date.The cause of the reported event was traced to component failure.
 
Manufacturer Narrative
Additional plant investigation information: an investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HEMODIALYSIS SYSTEM W/BIBAG
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer (Section G)
CONCORD MANUFACTURING
4040 nelson avenue
concord CA 94520
Manufacturer Contact
thomas c. johnson
920 winter st.
waltham, MA 02451
7816999499
MDR Report Key8151867
MDR Text Key129945468
Report Number2937457-2018-03642
Device Sequence Number1
Product Code KDI
UDI-Device Identifier00840861100910
UDI-Public00840861100910
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup,Followup
Report Date 12/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/11/2018
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number190766
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Device Age MO
Date Manufacturer Received12/21/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-