| Model Number |
100/399/218 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Premature Separation (4045)
|
| Health Effect - Clinical Code |
Awareness during Anaesthesia (1707)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
Information was received that a smiths medical portex lockit plus regional anesthesia catheter securement device became detached while in use with a patient.The reporter stated the detachment resulted in the catheter pulling out of the epidural space, which caused the epidural pain control to be insufficient.Subsequently, the nurses had to insert a new epidural catheter and a more invasive method of anesthesia intervention was required, noted as either, "general anesthesia or emergency cesarean section".No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One unused catheter securement device was returned for evaluation.Visual inspection of the device found the sample to be adhered to the fixture and the catheter was threaded through lockit.Functional testing of the device included setting the tensile tester to pull at 12 in per minute and stop at 6 in; noting that if catheter breaks or pulls, it is considered a failure.The sample remained in lockit for duration of the test, indicating an acceptable performance.The reported customer complaint was unable to be confirmed.This investigation revealed no intrinsic evidence to suggest a cause of issue related to manufacturing.
|
| |
|