(b)(6).Investigation summary: bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for color variation with the incident lot was observed.Additionally, retention samples were selected from bd inventory for evaluation and upon completion, the issue relating to color variation was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer photos, the customer¿s indicated failure mode for color variation with the incident lot was observed.Additionally, evaluation of the retain samples was conducted and color variation was not observed.Root cause description: based on the investigation, the most likely root cause was determined to be related to a manufacturing issue.As a result, bd is reviewing specific areas in the manufacturing process where the cause of this issue may have originated.
|
|
It was reported that a bd¿ tube media tube lihep plh had a cap with a lighter green color.The color was too light, similar to the grey cap."this makes the inpeco track the lihep tube no longer recognizes, places it in an error rack and analysis is postponed.This reduces the tat and the time to result, which results in delays in diagnosing patients.".
|