The investigation determined that a lower than expected vitros tsh result was obtained from a patient sample processed using vitros immunodiagnostic products tsh reagent in combination with a vitros 5600 integrated system.A definitive assignable cause for the event could not be determined.Based on historical quality control results, a vitros tsh reagent lot 5755 performance issue is not a likely contributor to the event.Additionally, continual tracking and trending of complaints has not identified any signals that would point to a potential systemic issue with vitros immunodiagnostic products tsh reagent lot 5755.Precision testing performed on the instrument yielded results that were within guidelines.Therefore, an instrument malfunction or unexpected performance is not a likely contributor to the event.In addition, pre-analytical sample processing could not be ruled out as a contributing factor, as it was established the customer was not following the sample collection device manufacturer¿s recommendation for sample centrifugation.It is possible that cellular debris, due to poor sample preparation, was present in the affected sample, although this could not be confirmed.Moreover, sample handling and storage along with sample mix up cannot be completely ruled out as possible contributing factors to the event as the exact testing date and time was unable to be confirmed.
|