• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX ULTRA TI 4.5 W/2 UB (WH & BLUE); FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX ULTRA TI 4.5 W/2 UB (WH & BLUE); FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 72202893
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
International zip code: (b)(6).
 
Event or Problem Description
It was reported that during insertion of the anchor into the patient's bone, the stem of the screwdriver has detached in its distal part, making it impossible to insert and remove the anchor from the bone.The surgeon removed the anchor using a large forcep.A backup device was available to complete the procedure with no delay or patient injuries.
 
Additional Manufacturer Narrative
One 72202893 twinfix ultra ti 4.5mm suture anchor reported on.Due to unavailability, the allegation could not be fully confirmed.Definitive conclusions, investigation and evaluation were not possible without physical evaluation.If objective evidence becomes available to assist with evaluation, the complaint will be revisited.Factors that may affect device performance include: device storage, device ability, surgical ability, procedure location and tissue condition.Final product met predetermined specifications upon release to distribution.
 
Event or Problem Description
It was reported that during insertion of the anchor into the patient's bone, the stem of the screwdriver detached in its distal part, making it impossible to insert and remove the anchor from the bone.Anchor was removed due to it was not completely inserted and the surgeon pull it out using a large forcep.The parts of the screwdriver were removed from patient since they are in the envelope.An additional hole had to be made and a backup device was available to complete the procedure with no delay or patient injuries.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA TI 4.5 W/2 UB (WH & BLUE)
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key8152473
Report Number1219602-2018-01724
Device Sequence Number11304207
Product Code MBI
UDI-Device Identifier03596010652287
UDI-Public03596010652287
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K100159
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number72202893
Device Lot Number2012761
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/29/2018
Supplement Date Received by Manufacturer01/10/2019
Initial Report FDA Received Date12/11/2018
Supplement Report FDA Received Date01/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-