Catalog Number 0064599
Medical Device Problem Code False Negative Result (1225)
Health Effect - Clinical Code Fever (1858)
Date of Event 11/21/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Immucor technical support used a remote electronic connection method to assess the testing instrument in question on 21nov2018, which showed test well images in question as visually negative.The review of the event log showed no errors during processing.The roi images appeared acceptable and the flatfields were acceptable.An immucor field service engineer (fse) visited the customer site to assess the testing instrument on (b)(6) 2018 and determined that the residual volume was slightly high out of specification with a value of 0.84.The fse adjusted that residual volume to within specification with a value of 0.52.The fse performed teach washer and the instrument subsequently passed qc.The internal immucor record number for this report is (b)(4).
 
Event or Problem Description
On (b)(6) 2018, a customer site reported an unexpectedly negative antibody screen when tested on a galileo neo instrument, which was involved in a transfusion reaction investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALILEO NEO
Common Device Name
AUTOMATED BLOOD BANK SYSTEM
Manufacturer (Section D)
IMMUCOR, INC.
3130 gateway drive
norcross GA 30091 5625
MDR Report Key8152528
Report Number1034569-2018-00279
Device Sequence Number11310117
Product Code KSZ
UDI-Device Identifier10888234001041
UDI-Public10888234001041
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
BK100033
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Catalogue Number0064599
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/21/2018
Initial Report FDA Received Date12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age66 YR
-
-