MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL IMPACTOR/POSTIONER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH UNIVERSAL IMPACTOR/POSTIONER; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 2101-0200
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of the device history records indicates that the reported devices were manufactured and accepted into final stock with no reported discrepancies.The complaint history review indicated that there were no similar events for the reported lot.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Not returned to the manufacturer.
 
Event or Problem Description
Threaded the trident 54 trial onto the acetbaular whittier impactor threaded piece was screwing past the cup trial.
 
Additional Manufacturer Narrative
Reported event: an event regarding assembly issue involving a cutting edge cup impactor was reported.The event was confirmed by product inspection.Method & results: -product evaluation and results: the dimensional inspection found that the protrusion was out of specification and examination by a material engineer indicated, "the plug welds on both ends were fractured.Partial press-fit was observed at the stud/handle interface." clinician review: not performed as medical records were not provided for review.Product history review: a device history review confirmed all devices accepted into finished goods conformed to specification.Complaint history review: no other events were reported for the lot indicated.Conclusions: the dimensional inspection confirmed the reported event.A material analysis engineer examined the device and noted that the plug welds were fractured on both ends.Nc and capa were previously opened to further investigate the protrusion of the threaded stud.See nc and capa for further details.No further investigation is needed at this time.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.Product surveillance will continue to monitor for trends.Corrective action/preventive action: nc and capa was previously raised to further investigate the protrusion of the threaded stud.The device has been redesigned and the drawing has been released on 15-mar-2018 under ecn and capa.
 
Event or Problem Description
Threaded the trident 54 trial onto the acetbaular whittier impactor threaded piece was screwing past the cup trial.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERSAL IMPACTOR/POSTIONER
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8152529
Report Number0002249697-2018-03980
Device Sequence Number16539004
Product Code LPH
UDI-Device Identifier07613327211511
UDI-Public07613327211511
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K161569
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number2101-0200
Device Lot Number134617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/20/2018
Initial Date Received by Manufacturer 11/13/2018
Supplement Date Received by Manufacturer01/18/2019
Initial Report FDA Received Date12/11/2018
Supplement Report FDA Received Date02/15/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age50 YR
Patient Weight96
-
-