BOSTON SCIENTIFIC CORPORATION XXL ESOPHAGEAL; DILATOR, ESOPHAGEAL Back to Search Results
Model Number 23375
Device Problems Material Rupture (1546); Device-Device Incompatibility (2919)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2018
Event Type  malfunction  
Event Description
It was reported that balloon rupture occurred.The target lesion was located in the right common vein.After a wallstent was implanted in the lesion, a 16-4/5.8/75cm xxl esophageal balloon catheter was advanced for post dilatation; however, during inflation at 2 atmospheres, the balloon ruptured.It was noted that the edge of the stent punctured the balloon.The device was removed and the procedure was completed with another of the same device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XXL ESOPHAGEAL
Type of Device
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8152649
MDR Text Key130046282
Report Number2134265-2018-63848
Device Sequence Number1
Product Code KNQ
UDI-Device Identifier08714729163060
UDI-Public08714729163060
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/20/2021
Device Model Number23375
Device Catalogue Number23375
Device Lot Number0022686112
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/27/2018
Initial Date FDA Received12/11/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-