• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS NIAGARA SLIM-CATH DIALYSIS CATHETER KIT 12F X 24CM (STRAIGHT) (SHORT-TERM) (D/L); CATHETER, HEMODIALYSIS, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS NIAGARA SLIM-CATH DIALYSIS CATHETER KIT 12F X 24CM (STRAIGHT) (SHORT-TERM) (D/L); CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 10/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a guidewire stuck in an introducer needle was confirmed and the cause was determined to be use related.The products returned for evaluation were one 0.038 in.J-tip guidewire in a plastic hoop and one 18 g introducer needle.The investigation findings are consistent with damage caused by retraction of the guidewire against the bevel of the introducer needle.The returned product sample was evaluated and the guidewire was confirmed to be broken with the coil wire being unraveled.It appeared that the inner core wire had broken which allowed the outer coil wire surrounding it to unravel.Microscopic examination of the fracture sites revealed the following: narrowing of the wire cross-section near the fracture site, which is a characteristic feature of a strong pull on the wire.Damage to the inside edge of the introducer needle which can occur if the guidewire is forcefully retracted against the needle, damaging the shape of the sharpened bevel biological material was also seen on the wire which may have contributed to the observed guidewire fracture as retraction of the wire together with the biological material could have caused the pieces to become stuck.Normal movement of the guidewire is away from the sharpened bevel and will not damage the wire.If the guidewire direction is reversed, the guidewire is then pulled against the sharpened edge of the needle bevel and can cause shearing damage of the wire and/or cause the wire to become stuck within the needle.The product ifu states: ¿do not pull back guidewire over needle bevel as this may sever the end of the guidewire.The introducer needle must be removed first.Also, if unusual resistance is met during manipulation of the guidewire, discontinue the procedure and determine the cause of resistance before proceeding.Withdraw needle and guidewire if cause of resistance cannot be determined.¿ an examination of the wire structure revealed no potential damage/defect related to manufacture of the product.A lot history review (lhr) of recp1968 showed no other similar product complaint(s) from this lot number.
 
Event or Problem Description
It was reported that the guide wire was stuck at the position of the needle and could not be pulled out.On (b)(6) 2018, returned wire is broken.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIAGARA SLIM-CATH DIALYSIS CATHETER KIT 12F X 24CM (STRAIGHT) (SHORT-TERM) (D/L)
Common Device Name
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key8152736
Report Number3006260740-2018-03556
Device Sequence Number2325156
Product Code MPB
UDI-Device Identifier00801741045370
UDI-Public(01)00801741045370
Combination Product (Y/N)N
PMA/510(K) Number
K010778
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number5553240
Device Lot NumberRECP1968
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2018
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2018
Initial Report FDA Received Date12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age81 YR
-
-