STRYKER ORTHOPAEDICS-MAHWAH PRIMARY TRITANIUM HEMI SOLIDBACK CUP 56MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
|
Back to Search Results |
|
| Catalog Number |
500-03-56E |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Codes |
Pain (1994); Injury (2348); Inadequate Osseointegration (2646)
|
| Date of Event |
11/13/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding loosening and pain involving a tritanium shell was reported.The event was not confirmed.Method & results: product evaluation and results: not performed as the device was not returned.Medical records received and evaluation: not performed as no medical records were returned for review.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there were no other events for the lot provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as device return, pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time as no devices and insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.Device not returned.
|
| |
|
Event or Problem Description
|
|
It was reported that patient's left hip was revised due to pain and aseptic loosening of the shell.A stryker shell, liner, and head were revised to a tritanium revision shell, all-poly liner, and ceramic head.
|
| |
|
Search Alerts/Recalls
|
|
|