| Model Number |
MSERIES BI-PHASIC |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
Death (1802)
|
| Date of Event |
11/20/2018
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to defibrillate a male patient (age unknown), the device failed to discharge.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.Complainant indicated that during subsequent biomed testing, the malfunction was not duplicated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, defib cycling, shock testing and internal inspections without duplicating the report.The device was recertified and returned to the customer.The electrode pads used during the event were not returned as part of this investigation.No clinical data was available.No trend is associated with reports of this type.
|
| |
|