Model Number MSERIES BI-PHASIC
Medical Device Problem Code Failure to Deliver Shock/Stimulation (1133)
Health Effect - Clinical Code Death (1802)
Date of Event 11/20/2018
Type of Reportable Event Death
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that while attempting to defibrillate a male patient (age unknown), the device failed to discharge.Complainant indicated that the clinician obtained another device to continue treating the patient.Complainant indicated that the patient subsequently expired.Complainant indicated that during subsequent biomed testing, the malfunction was not duplicated.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation and the customer's report was not replicated or confirmed.The device was put through extensive testing including bench handling, defib cycling, shock testing and internal inspections without duplicating the report.The device was recertified and returned to the customer.The electrode pads used during the event were not returned as part of this investigation.No clinical data was available.No trend is associated with reports of this type.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key8152837
Report Number1220908-2018-03561
Device Sequence Number14000880
Product Code MKJ
UDI-Device Identifier00847946013738
UDI-Public00847946013738
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K990762
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Other
Device Model NumberMSERIES BI-PHASIC
Device Catalogue NumberM SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/21/2018
Supplement Date Received by Manufacturer11/21/2018
Initial Report FDA Received Date12/11/2018
Supplement Report FDA Received Date12/24/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death;
-
-