| Catalog Number |
60-7510-005 |
| Medical Device Problem Codes |
Misfire (2532); Firing Problem (4011)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Inspection was unable to find any abnormalities or defects.The device was functionally inspected using a system 7500 esu ((b)(4)).Functional inspection found that both the cut and coag modes functioned as expected.The buttons did not require any extra effort or pressure to activate.No auto-activation of the device was observed during testing.Upon dismantling the device no issues were observed with the buttons, circuit board, or dome switch.A lot history review was conducted, and this is the only complaint for this lot number and failure code.The manufacturing document from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution were found to have met all specifications prior to shipment.A two-year review of complaint history revealed there has been a total of 7 complaints, regarding 7 devices for this device family and failure mode.During the same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed, the rate of failure would be (b)(4).Per the instructions for use, the user is advised the following; - these devices should be inspected before each use.This issue will continue to be monitored through the complaint system to assure patient safety.
|
| |
|
Event or Problem Description
|
|
The customer reported that the (b)(4), goldvac pb smoke pencil, while starting the circumcision, surgical procedure on (b)(6) 2018 the electrosurgical smoke evacuation handpiece fired without surgeon touching button.The surgeon was about to start using handpiece, aimed in the direction of the surgical location.Device button was not pressed yet according to surgeon the device engaged in cut mode.Surgeon doubled checked his hand placement.Device did go off on its own again.The cut on the penis was where the surgeon was going to initiate the surgical incision, so according to the doctor there were no issues an no harm to the patient.The yellow button on handpiece was difficult to press and appeared stuck and difficult to release.This report is being raised based on device malfunction with potential for injury upon reoccurrence.
|
| |
|
Search Alerts/Recalls
|