• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP MONOBLOCK; RSP, REVISION HUMERAL MONOBLOCK, 12MM X 175

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP MONOBLOCK; RSP, REVISION HUMERAL MONOBLOCK, 12MM X 175 Back to Search Results
Model Number 510-12-175
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/26/2018
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery: the stem began to loosen.
 
Additional Manufacturer Narrative
The reason for this revision surgery was due to loose stem.The previous surgery and the revision surgery detailed in this investigation occurred 3.7 years apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.This investigation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - (b)(4) for examination.A review of the implant device history records (dhr), shows that the reported component(s) used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmrs) associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of use during the previous surgery.Customer complaint history of the reported device(s) showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to loose stem.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP MONOBLOCK
Common Device Name
RSP, REVISION HUMERAL MONOBLOCK, 12MM X 175
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
MDR Report Key8153193
Report Number1644408-2018-01116
Device Sequence Number2364170
Product Code KWS
UDI-Device Identifier00888912144766
UDI-Public(01)00888912144766
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K100741
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/24/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/28/2019
Device Model Number510-12-175
Device Catalogue Number510-12-175
Device Lot Number986F1016
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/26/2018
Supplement Date Received by Manufacturer01/18/2019
Initial Report FDA Received Date12/11/2018
Supplement Report FDA Received Date01/24/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
508-01-436, LOT 923F1035; 508-36-101, LOT 869C1186
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-