Model Number 6935M72
Medical Device Problem Codes Over-Sensing (1438); Pacing Problem (1439); Decreased Sensitivity (2534)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/02/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Concomitant medical products: dtmb1qq crt-d, implanted: (b)(6) 2017.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that an in-office check found that the right ventricular (rv) lead had very small r-waves.There had also been significant episodes of t-wave oversensing (twos) earlier in the month of which the t-wave discrimination feature applied and kept the patient from getting inappropriate shocks.It was noted that due to the t-wave oversensing post vs (ventricular sensed) there was a loss of crt (cardiac resynchronization therapy) pacing.The rv lead was reprogrammed and remains in use.No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO SECURE S
Common Device Name
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key8153473
Report Number2649622-2018-20418
Device Sequence Number252418
Product Code LWS
UDI-Device Identifier00613994718877
UDI-Public00613994718877
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/11/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/28/2015
Device Model Number6935M72
Device Catalogue Number6935M72
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2018
Initial Report FDA Received Date12/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
Patient Weight87
-
-