| Model Number |
406844 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/20/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the procedure, blood was leaking from the introducer.The introducer was replaced and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Event or Problem Description
|
|
During an atrial fibrillation procedure, blood was leaking from the introducer.The introducer was replaced and the procedure was completed with no adverse consequences to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of leak could not be confirmed.The sheath passed pressure and aspiration leak testing with no anomalies noted.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported leak remains unknown.
|
| |
|