| Model Number |
1032800 |
| Medical Device Problem Code |
Temperature Problem (3022)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/15/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The manufacturer received information alleging a ventilator was alarming for a high temperature alarm condition.There was no harm or injury reported.The device has yet to be returned to the manufacturer for evaluation.At this time, we are unable to confirm the alleged malfunction.A follow-up report will be submitted when the manufacturer's investigation is complete.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously reported a ventilator was alarming for a high temperature alarm condition.There was no harm or injury reported.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The increased airstream temperature (high temperature alarm) appears to have been caused by the ambient conditions in which the device was operating.No malfunctions were noted.The device was found to audibly and visually alarm, as designed.
|
| |
|