| Catalog Number |
01.26.45.0058 |
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
| Date of Event |
11/13/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Clinical evaluation performed by medical affairs manager: partial hip revision surgery (head, cup and liner) occurred 7 months after primary cementless total hip arthroplasty.Radiographic image provided shows acetabular component mobilization.Primary cup position cannot be assessed as no immediate postoperative x-ray is available.No conclusion can be drawn on the basis of information received.Batch review performed on 11 december 2018: lot 134533: (b)(4) items manufactured and released on 22-nov-2013.Expiration date: 2018-10-31.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any similar reported event.
|
| |
|
Event or Problem Description
|
|
The patient came in due to instability 7 months after primary which was caused by the cup becoming loose inside of the acetabulum.The surgeon revised the head, cup and liner and the surgery was completed successfully.
|
| |
|