• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION R SERIES DEFIBRILLATOR; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number R SERIES
Medical Device Problem Code Unable to Obtain Readings (1516)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
 
Event or Problem Description
Complainant alleged that during functional testing, the device displayed a "poor pad contact" message.Complainant indicated that there was no patient involvement in the reported malfunction.
 
Additional Manufacturer Narrative
The reported malfunction was observed and attributed to a faulty test port, pins worn out.This is a malfunction which only impacts self-test portion with the defibrillator.Once the multi-function connection is removed from the test port and attached to a set of electrode pads or external paddles the device performs to specification.This type of malfunction does not pose any clinical impact and does not meet our guidelines for reportability.Analysis of reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R SERIES DEFIBRILLATOR
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key8155250
Report Number1220908-2018-03589
Device Sequence Number16551803
Product Code DRT
UDI-Device Identifier00847946017194
UDI-Public00847946017194
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K060559/P160
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/21/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberR SERIES
Device Catalogue NumberR SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 11/21/2018
Supplement Date Received by Manufacturer11/21/2018
Initial Report FDA Received Date12/12/2018
Supplement Report FDA Received Date01/14/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-