• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. BD VACUTAINER® SODIUM HEPARIN¿ (NH) BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. BD VACUTAINER® SODIUM HEPARIN¿ (NH) BLOOD COLLECTION TUBES; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 366480
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/27/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event or Problem Description
It was reported that bd vacutainer® sodium heparin (nh) blood collection tubes had a milky white powder with a film over it stuck at the bottom of the tubes.
 
Event or Problem Description
It was reported that bd vacutainer® sodium heparin¿ (nh) blood collection tubes had a milky white powder with a film over it stuck at the bottom of the tubes.
 
Additional Manufacturer Narrative
Investigation summary: bd received samples and photos from the customer facility for investigation.The photos were evaluated and the customer¿s indicated failure mode for foreign matter with the incident lot was not observed.Additionally, evaluation of the customer samples was performed and foreign matter was not observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Investigation conclusion: based on evaluation of the customer samples, the customer¿s indicated failure mode for foreign matter with the incident lot was not observed as all samples met the required specifications.Additionally, the photos did not identify a specific product issue.Root cause description: based on the investigation, a root cause could not be determined.The product was found to be in conformance and meet release specifications.Rationale: complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.The bd business team regularly reviews the collected data for identification of emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® SODIUM HEPARIN¿ (NH) BLOOD COLLECTION TUBES
Common Device Name
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key8156038
Report Number1917413-2018-03970
Device Sequence Number1903038
Product Code JKA
UDI-Device Identifier50382903664806
UDI-Public50382903664806
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
N/A
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2020
Device Catalogue Number366480
Device Lot Number8099579
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2018
Initial Date Received by Manufacturer 11/27/2018
Supplement Date Received by Manufacturer11/27/2018
Initial Report FDA Received Date12/12/2018
Supplement Report FDA Received Date01/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-