• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. 1818910 S-ROM*STM STD,36NK,14X09X130; S-ROM HIP SYSTEM : HIP FEMORAL STEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. 1818910 S-ROM*STM STD,36NK,14X09X130; S-ROM HIP SYSTEM : HIP FEMORAL STEM Back to Search Results
Catalog Number 523191
Medical Device Problem Codes Loose or Intermittent Connection (1371); Migration (4003)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646); No Code Available (3191)
Date of Event 11/16/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
The patient was revised due to hip pain.X-rays shows subsidence of the sleeve and stem.The doctor revised the femoral component.Cup and poly as well.Stem and sleeve loosening at bone to implant interface.Doi: (b)(6) 2013, dor:nov 16, 2018, right hip.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).Investigation summary: examination of the returned devices was not able to confirm the reported event.No evidence was found indicating product error was a contributing factor.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch , a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic drive
warsaw IN 46582 0988
MDR Report Key8156080
Report Number1818910-2018-77917
Device Sequence Number11195529
Product Code JDI
UDI-Device Identifier10603295170853
UDI-Public10603295170853
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K851422
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number523191
Device Lot Number2507935
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2018
Initial Date Received by Manufacturer 11/16/2018
Supplement Date Received by Manufacturer01/07/2019
Initial Report FDA Received Date12/12/2018
Supplement Report FDA Received Date01/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
-
-