• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH 21MM MOD REV HIP BDY/BLT +20MMCOMPONENT LEVEL 9006-1-221; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH 21MM MOD REV HIP BDY/BLT +20MMCOMPONENT LEVEL 9006-1-221; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 6276-1-221
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code Injury (2348)
Date of Event 11/14/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
An event regarding limb length discrepancy involving a restoration modular body was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Medical records received and evaluation: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.No further investigation is required at this time.The exact cause of the event could not be determined because insufficient information was provided.Additional information, including operative reports, progress notes, x-rays and return of the device "are" needed to fully investigate the event.If further information becomes available and/or the product is returned, this investigation will be re-opened.Product surveillance will continue to monitor for trends.
 
Event or Problem Description
It was reported that patient's right hip was revised.As reported by rep: "patient appeared to have offset issues on x-ray".A 21mm +20 v40 restoration proximal body and 36 +0 biolox v40 head were revised to a 25mm +20 v40 restoration proximal body and another 36 +0 biolox v40 head.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
21MM MOD REV HIP BDY/BLT +20MMCOMPONENT LEVEL 9006-1-221
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8156282
Report Number0002249697-2018-04007
Device Sequence Number267786
Product Code LPH
UDI-Device Identifier04546540641212
UDI-Public04546540641212
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2020
Device Catalogue Number6276-1-221
Device Lot Number52635205
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2018
Initial Report FDA Received Date12/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age83 YR
-
-