MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION LIGHTTRAIL SINGLE USE LASER FIBER; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION LIGHTTRAIL SINGLE USE LASER FIBER; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M006864640
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplement report will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that a lighttrail 270um single use laser fiber was opened for use in a flexible ureteroscopy procedure in the kidney performed on (b)(6) 2018.According to the complainant, during preparation, the laser body broke upon removal from the package.Reportedly, there was no damage noted to the package or device prior to opening the product pouch.The procedure was completed with another laser fiber.There were no adverse patient effects reported as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation on november 16, 2018 that a lighttrail 270um single use laser fiber was opened for use in a flexible ureteroscopy procedure in the kidney performed on (b)(6) 2018.According to the complainant, during preparation, the laser body broke upon removal from the package.Reportedly, there was no damage noted to the package or device prior to opening the product pouch.The procedure was completed with another laser fiber.There were no adverse patient effects reported as a result of this event.
 
Additional Manufacturer Narrative
Problem code 1069 captures the reportable event of fiber broke.A lighttrail single use laser fiber was returned broken.The fiber measured approximately 201.1 cm from the top of the connector to the break.The remainder of the fiber, including the distal tip, was not returned.Review and analysis of all available information indicated that the damage to the device is likely a result of handling.Therefore the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and no deviations were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIGHTTRAIL SINGLE USE LASER FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
model farm road,
marlborough MA 01752
MDR Report Key8156292
Report Number3005099803-2018-61900
Device Sequence Number16685712
Product Code GEX
UDI-Device Identifier08714729894995
UDI-Public08714729894995
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K111475
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/06/2022
Device Model NumberM006864640
Device Catalogue Number6464
Device Lot Number0201708165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/03/2018
Initial Date Received by Manufacturer 11/16/2018
Supplement Date Received by Manufacturer01/17/2019
Initial Report FDA Received Date12/12/2018
Supplement Report FDA Received Date02/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-