COOK ENDOSCOPY TRACER HYBRID WIRE GUIDE; OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY Back to Search Results
Catalog Number HYB-48015
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: cook tri-ex extraction balloon, tx-8.5-a investigation evaluation: our evaluation of the product said to be involved confirmed the report.There is wire guide coating damage near the distal end.The coil spring is still attached to the distal end of the wire guide.The third, fifth, and sixth centimeter markings appear stretched.Seven (7) mm of the distal core wire is exposed.Approximately 3.1 cm of coating is loose from the core wire and hanging off the wire guide just proximal to the bare section.The coil spring is protruding in two (2) places from that loose portion of wire guide coating.A few areas feel slightly rough throughout the length of the wire guide.A product-specific discrepancy that could have caused or contributed to this observation was not observed during our laboratory analysis.A review of the wire guide subassembly device history records were performed.A discrepancy or anomaly was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.The instructions for use instructs the user to do the following: "flush endoscope accessory channel and/or lumen of device with sterile water, then insert wire guide floppy end first.Note: for best results, wire guide should be kept wet, if applicable." failure to flush the endoscope channel of device lumen can result in damage to the wire guide.If additional pressure is applied to the wire guide and/or accessory device(s) while moving the wire guide inside the accessory device(s), this could contribute to wire guide coating damage.Prior to distribution, all tracer hybrid wire guides are subjected to a visual inspection to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook tracer hybrid wire guide.The coating stripped off of the tip of the device.It did not detach and was removed with the rest of the device.The procedure was successfully completed with another manufacturer's wire guide.Additional information was received on 26-nov-2018: the damage is at the tip of the device.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRACER HYBRID WIRE GUIDE
Common Device Name
OCY ENDOSCOPIC GUIDEWIRE, GASTROENTEROLOGY-UROLOGY
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key8156619
Report Number1037905-2018-00584
Device Sequence Number11241047
Product Code OCY
UDI-Device Identifier00827002226593
UDI-Public(01)00827002226593(17)210917(10)W4115705
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 11/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/17/2021
Device Catalogue NumberHYB-48015
Device Lot NumberW4115705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/26/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2018
Initial Report FDA Received Date12/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-