• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROVASCULAR AXIUM PRIME BARE PL 3D; DEVICE, EMBOLIZATION, VASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROVASCULAR AXIUM PRIME BARE PL 3D; DEVICE, EMBOLIZATION, VASCULAR Back to Search Results
Model Number APB-1.5-4-3D-ES
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/04/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received a report that, when the instant detacher was pulled out to detach the coil, the coil came out together still attached.Multiple attempts were made using the instant detacher without resolve, so a manual detachment method was performed, resulting in the coil finally detaching.It was noted that the physician has had many detachment errors with the coil type.The patient was undergoing embolization surgery for treatment of a 3mm aneurysm.It was noted that the patient had a slightly severe vessel tortuosity and a slightly small vessel tortuosity.There were no related patient symptoms.Ancillary devices include fubuki 6fr, shouryu4*7, echelon 10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AXIUM PRIME BARE PL 3D
Common Device Name
DEVICE, EMBOLIZATION, VASCULAR
Manufacturer (Section D)
MEDTRONIC NEUROVASCULAR
9775 toledo way
irvine CA 92618
MDR Report Key8156633
Report Number2029214-2018-01054
Device Sequence Number13841114
Product Code KRD
UDI-Device Identifier00847536021167
UDI-Public00847536021167
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K081465
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2021
Device Model NumberAPB-1.5-4-3D-ES
Device Catalogue NumberAPB-1.5-4-3D-ES
Device Lot NumberA597498
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2018
Initial Report FDA Received Date12/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age57 YR
-
-