| Model Number |
10604 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
10/05/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device is a combination product.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 19-dec-2018.It was reported that difficulty loading wire occurred.During introduction of a 2.25 x 24mm synergy ii drug-eluting stent into the wire, the device would not go over the wire.Another 2.25 x 24mm synergy stent was used and the procedure was completed.No patient complications or injuries were reported.However, returned device analysis revealed a stent damage.A synergy ii us mr 2.25 x 24mm stent delivery system (sds) was returned for analysis without a stent.A visual and microscopic examination of the crimped identified stent damage.The crimped stent outer diameter (od) of the undamaged section of the crimped stent was measured and is within max crimped stent profile measurement.The balloon was reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and tactile examination of the hypotube found multiple hypotube kinks.A visual and tactile examination of the shaft polymer extrusion revealed no issues.The device was loaded onto and tracked over a 0.014 guidewire without issue.A visual and microscopic examination found damage to the tip.No other issues were identified during the product analysis.
|
| |
|