| Model Number |
130003V |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
Information was received that a smiths medical pneupac ventipac medical ventilator had a damaged gauge reading.No adverse effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation results: one pneupac ventilator was received device in good condition.The manometer was not physically centered.Also, the gaugeneedle resting position was outside the +/- 2 cmh2o tolerance.These observations confirmed the customer's complaint.
|
| |
|