Product complaint # (b)(4).Date sent to the fda: 1/22/2019.Evaluation: representative samples returned to support investigation of multiple device issues captured in reports 2210968-2018-77720, 2210968-2018-77721, 2210968-2018-77722, 2210968-2018-77723, and 2210968-2018-77724.Analysis results have been included in follow-up reports for each.An unopened sample was returned for evaluation.The issued sample was not received.During the visual inspection of the sample, no defects were found on the package.The sample was opened, and the swage and attachment area were noted to be as expected.The suture was dispensed without problems and examined along the strand with no defects or damage observed.A functional test was performed and the pull force was above the minimum requirements.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the conditions of the sample received, no suture needle pull off was found, and the tested sample met the finished goods requirements.
|