| Model Number |
NS2007 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Uterine Perforation (2121)
|
| Date of Event |
11/15/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device is not being returned therefore, a failure analysis of the complaint device cannot be completed.Device history record (dhr) review was conducted for the reported identification number.The lot was released meeting all qa specifications.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a novasure endometrial ablation, a hysteroscopy was performed and then the physician attempted to seat the novasure device.The physician was only able to achieve a width of 2cm and "repeated seating attempts a few times before moving to a second device." upon using a second device the width was unchanged at 2cm.One last attempt was made and the second device opened to 4.5cm.The physician suspected a perforation and inspected the uterine cavity with a hysteroscope and confirmed a perforation.The procedure was aborted.No additional information was received.
|
| |
|
Search Alerts/Recalls
|