| Model Number |
MT22608 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(4) 2018 that on (b)(6) 2018, the receiver had intermittent audio output.It was reported that an adult patient was using a pediatric receiver.Labeling indicates: the dexcom g4 platinum (pediatric) continuous glucose monitoring system is a glucose monitoring device indicated for detecting trends and tracking patterns in persons ages 2 to 17 years with diabetes.No product was provided for evaluation.The complaint confirmation of the receiver having intermittent audio output could not be determined.A probable cause could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The receiver was returned for evaluation.An exterior visual inspection was performed and passed.The receiver was charged and will boot.The receiver log was downloaded and reviewed, finding no errors related to the complaint.A receiver functional test was performed and passed all relevant testing.A global communication tool software test was performed and passed.Wiggle testing was performed and passed.Manual "try it" testing was performed and passed.The receiver case was opened for an interior inspection and passed.Speaker resistance was measured and was within specification.The reported event of an intermittent audio output was not confirmed.A probable cause could not be determined.
|
| |
|