• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number MICL13.2
Medical Device Problem Code Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/12/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
The reporter indicated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -11.0 diopter, in the patients right eye (od) on (b)(6) 2018.The lens was explanted and exchanged within the same surgery, on (b)(6) 2018 due to excessive vaulting.The lens was exchanged for a shorter lens and the problem was resolved.
 
Additional Manufacturer Narrative
Device evaluated by mfr : device evaluation: the lens was returned in liquid in lens vial.Visual inspection found one haptic torn.There was evidence of residue on the lens.Corrected data: date of event -(b)(6)2018 claim #(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Common Device Name
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia CA 91016
MDR Report Key8158090
Report Number2023826-2018-01829
Device Sequence Number11134709
Product Code MTA
UDI-Device Identifier00841542103312
UDI-Public00841542103312
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P030016
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2019
Device Model NumberMICL13.2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2018
Initial Date Received by Manufacturer 11/12/2018
Supplement Date Received by Manufacturer12/18/2018
Initial Report FDA Received Date12/12/2018
Supplement Report FDA Received Date01/18/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CARTRIDGE MODEL AND LOT NUMBER UNK; INJECTOR MODEL AND LOT NUMBER UNK
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
-
-