| Model Number |
MICL13.2 |
| Medical Device Problem Code |
Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm micl13.2 implantable collamer lens, -12.0 diopter, in the patients left eye (os) on (b)(6) 2018.The lens was explanted and exchanged within the same surgery, on (b)(6) 2018 due to excessive vaulting.The lens was exchanged for a shorter lens and the problem was resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
The lens was returned in liquid in a lens vial.Visual inspection found one haptic torn.Date of event: (b)(6) 2018.(b)(4).
|
| |
|
Search Alerts/Recalls
|