| Model Number |
ZCT450 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/29/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: although an exact date was not provided, the initial report stated the rotation was noted one week post implant, thus (b)(6) 2018 is provided as a best estimate.If explanted; give date: not applicable as to date the lens remains implanted and therefore not explanted.(b)(4).Device evaluation: product testing could not be performed because the product was not returned as to date it remains implanted.A product quality deficiency could not be confirmed.Manufacturing records review: during the manufacturing record review of the production order for this serial number, no product deficiency was identified.The documentation shows that the production order was manufactured according to specifications.Historical data analysis: a search of complaints revealed no similar complaints were received for this production order number.Labeling review: the labeling review was completed and revealed that the directions for use (dfu) provide the customer with proper usage instructions and guidelines.No labeling changes are required.Conclusion: as a result, of the investigation there is no indication of a product malfunction or product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the zct450 14.5 diopter intraocular lens (iol) was implanted in the patient¿s right eye (od) on (b)(6) 2018.One week post-op it was reported the lens rotated 10 degrees counter-clockwise and visual acuity was 20/70.Through follow up, customer confirmed that the patient did not have any health related issue and/or medications that could have contributed to the zct450 rotation.Although there was minimal rotation counter clockwise again, the patient has elected to wear prescription glasses because it corrects to 20/20 od with refraction of -1.00 +1.25 x 045.No additional information was provided to johnson & johnson surgical vision.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|
Search Alerts/Recalls
|