| Model Number |
G34279 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
510 (k) number; k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
As reported to customer relations "the device was placed down the scope and the needle would not exit the sheath.The device was removed from the scope and another of the same device was used to complete the procedure.No patient care was impacted." update (b)(6) 2018- per physician - not sure if what we saw was a broken piece.Could just be a clot.I looked after we saw during bronchoscopy and couldn¿t find anything.
|
| |
|
Event or Problem Description
|
|
This report is being submitted as a cancellation report based on the evaluation of the device involved and receipt of information "confirming" that no section of the device detached inside the patient¿s body.Initial report details: as reported to customer relations "the device was placed down the scope and the needle would not exit the sheath.The device was removed from the scope and another of the same device was used to complete the procedure.No patient care was impacted." update 16nov2018 - per physician - not sure if what we saw was a broken piece.Could just be a clot.I looked after we saw during bronchoscopy and couldn¿t find anything.
|
| |
|
Additional Manufacturer Narrative
|
|
510 (k) number; k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4).Cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: (b)(4).This report is being submitted as a cancellation report based on the evaluation of the device involved and receipt of information "confirming" that no section of the device detached inside the patient¿s body.Investigation details: two devices related to this occurrence underwent a laboratory evaluation.The needle was found to be kinked distally on the first device.Another kink proximal to the sheath extender was also observed on the first device however it was determined that this proximal kink was not observed prior to shipping back to cirl and as a result could be assumed to be due to transport returns.No defect was observed on the second device.Prior to distribution, all echo-hd-22-ebus-p-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo-hd-22-ebus-p-c of lot number: c1530530 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number: c1530530.The notes section of the instructions for use, which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.The failure of needle kinked/bent was concluded from the available information.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to the device being used in a flexed or twisted position during the procedure.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|
Additional Manufacturer Narrative
|
|
510 (k) number; k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4).Cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is still pending.A follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
As reported to customer relations "the device was placed down the scope and the needle would not exit the sheath.The device was removed from the scope and another of the same device was used to complete the procedure.No patient care was impacted." update 16nov2018 - per physician - not sure if what we saw was a broken piece.Could just be a clot.I looked after we saw during bronchoscopy and couldn¿t find anything.
|
| |
|