|
This follow-up report is being filled to relay investigation result.Dhr review: the device manufacturing quality records indicate that the released components met all requirements to perform as intended.Trend analysis: no trend considering the following event is identified: dislocation.Event description: it was reported that a 63 years old female patient, bmi=28.7, underwent initial surgery on (b)(6) 2017 and revised on (b)(6) 2018 due to dislocation.Review of received data: assessment of imaging (see the detailed x-ray review report in (b)(4) - warsaw split case): ap pelvis and 2 views of the left hip dated (b)(6) 2018 demonstrate bilateral total hip arthroplasties.There is screw fixation of the left acetabular cup.Symmetric position of the left femoral head within the acetabular cup.Neutral position of the femoral stem within the femur.No radiolucency to suggest loosening.No fracture.Acetabular inclination angle on the ap pelvis is 49.2.Unable to assess acetabular inclination angle on the crosstable lateral film.No abnormalities concerning the ceramic head is visible.Review of operative notes dated (b)(6) 2017 confirm that the patient underwent an initial left total hip arthroplasty.The acetabular shell was impacted using an intraoperative alignment jig at 45 degrees abduction and 20 degrees anteversion.Excellent rim fit was obtained.The femur was sequentially rasped to excellent fit and stability.The hip was reduced and stability was noted through range of motion.The final implants were impacted and limn length was verified.No complications were noted.Review of medical note dated 27 nov 2018 confirmed that patient received an exam for left hip instability as well as multiple dislocations (no dates given) while performing normal tasks.The surgeon further contributes this recurrence on the patient's diagnosis of ankylosing spondilitis, an inflammatory condition of the spine that causes scarring and immobility between the vertebrae, ultimately affecting gait and posture which can lead to dislocations.The surgeon discussed the plan for revision of the head and liner with a constrained liner.No revision surgery report was received.However, the po request form indicates the implantation of a biolox option head with a constrained liner for the surgery.Device analysis: no product was returned to zimmer biomet for in-depth analysis.Review of product documentation: this device is intended for treatment.The compatibility check was performed from www.Productcompatibility.Zimmer.Com and showed that the product combination was approved by zimmer biomet.Instruction for use (ifu) is reviewed.Conclusion: patient underwent revision surgery due to dislocation after 1 year 6 months in-vivo time.The quality records show that all specified characteristics (material, dimensions, surface, etc.) have met the specifications valid at the time of production.The investigation results did not identify a non-conformance or a complaint out of box (coob).Medical notes confirm the reported dislocation event.The most likely root cause for the event is the patient's diagnosis of ankylosing spondilitis, an inflammatory condition of the spine that causes scarring and immobility between the vertebrae, ultimately affecting gait and posture which can lead to dislocations.However, an exact root cause cannot be defined with the information available at the hand.The need for corrective measures is not indicated and zimmer gmbh considers this case as closed.Zimmer biomet¿s reference number of this file is (b)(4).
|