Model Number EX1150
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Information (3190)
Event Date 12/06/2018
Event Type  malfunction  
Manufacturer Narrative
No complaint was received with the return of the device. Failure event observed during analysis. Final analysis found burnt components on the main pcb and inner casings of the alternating current (ac) power adapter. Burnt modem circuit components were observed on the main pcb. The damage was sustained through with a high current flow through the ac electrical power outlet.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameMERLIN@HOME, RF TELEMETRY BASIC
Type of DevicePACEMAKER DATA TRANSMITTER
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key8160108
MDR Text Key130206032
Report Number2017865-2018-18838
Device Sequence Number1
Product Code DRG
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial
Report Date 12/13/2018
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received12/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberEX1150
Device Catalogue NumberEX1150
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2017
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received12/06/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-