• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0° X3 INSERT 32MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENT 0° X3 INSERT 32MM ID; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Catalog Number 623-00-32G
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/15/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.Review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.There have been no other events for the lot referenced.Not available.
 
Event or Problem Description
Customer reported that the liner would not seat into the acetabular shell and would come away with the impactor.Stryker rep reported that the case is just happening, and the customer got another implant (same product code) and it has seated in the shell.There was a delay of approximately 15 minutes whilst attempting to seat the prosthesis and then proceeded to get another implant.
 
Additional Manufacturer Narrative
An event regarding seating/locking issues involving a trident liner was reported.The event was not confirmed.Method & results: device evaluation and results: the returned device was examined by an engineer who noted the following: damage observed on insert consistent with attempted implantation.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Medical records received and evaluation: could not be performed as no medical records were returned for review.Product history review: a review of the device manufacturing records indicate that all devices accepted into final stock were free from discrepancies.Complaint history review: there were no other events for the lot provided.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Additional information including post operative x-rays and the primary operative report are needed to fully investigate the event.If further information becomes available, this investigation will be re-opened.
 
Event or Problem Description
Customer reported that the liner would not seat into the acetabular shell and would come away with the impactor.Stryker rep reported that the case is just happening, and the customer got another implant (same product code) and it has seated in the shell.There was a delay of approximately 15 minutes whilst attempting to seat the prosthesis and then proceeded to get another implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT 0° X3 INSERT 32MM ID
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key8160195
Report Number0002249697-2018-04030
Device Sequence Number11313564
Product Code LPH
UDI-Device Identifier07613327039481
UDI-Public07613327039481
Combination Product (Y/N)N
Initial Reporter CountryGB
PMA/510(K) Number
K033716
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Catalogue Number623-00-32G
Device Lot Number59639001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/15/2019
Initial Date Received by Manufacturer 11/15/2018
Supplement Date Received by Manufacturer03/20/2019
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date04/17/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-