Catalog Number 4002C1245
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/14/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the left internal iliac artery (iia) using penumbra coil 400s (pc400s).During the procedure, the physician felt resistance and was unable to advance a pc400 within its introducer sheath.Therefore, the pc400 was removed.The procedure was completed using a new pc400.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was intact at the proximal end of the pusher assembly.The pusher assembly was undamaged.The embolization coil was undamaged and intact with the pusher assembly distal detachment tip (ddt).Conclusion: evaluation of the returned pc400 revealed an undamaged, fully functional device.The device was able to advance in and out of the introducer sheath multiple times without issue.The device was also able to advance through a demonstration microcatheter without issue.The reported complaint could not be confirmed.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key8160240
Report Number3005168196-2018-02433
Device Sequence Number1825878
Product Code HCG
UDI-Device Identifier00814548010649
UDI-Public00814548010649
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/15/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/11/2025
Device Catalogue Number4002C1245
Device Lot NumberF77629
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/15/2018
Supplement Date Received by Manufacturer02/08/2019
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date03/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-