BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7010
Device Problem Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/12/2018
Event Type  malfunction  
Event Description
Reportable based on device analysis completed on 21-nov-2018.It was reported that shaft kink occurred.The target lesion was located in a coronary artery.A 3.5mm x 12mm quantum maverick balloon catheter was advanced for treatment.However, it was noted that the catheter was kinked.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.However, returned device analysis revealed shaft detached/separated.Device evaluated by manufacturer: the returned product consisted of the proximal portion of the quantum maverick balloon catheter.Only the hub and 65.3cm of the hypotube were returned for analysis.Portion of the hypotube, the distal shaft, balloon, markerbands and tip were not returned for analysis.The hypotube was visually examined.There were numerous hypotube kinks throughout the returned portion of the returned device.There was a complete hypotube separation 65.3cm distal of the strain relief.The hypotube fracture surface was ovaled, which suggests the device was kinked prior to separation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC SCIMED, INC
two scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
two scimed place
maple grove, MN 55311
6515827403
MDR Report Key8160327
MDR Text Key130225195
Report Number2134265-2018-64021
Device Sequence Number1
Product Code LOX
UDI-Device Identifier08714729392699
UDI-Public08714729392699
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P860019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Physician
Type of Report Initial
Report Date 12/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/28/2021
Device Model Number7010
Device Catalogue Number7010
Device Lot Number0022179220
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/05/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/21/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-