| Model Number |
DL-344-10-K |
| Medical Device Problem Codes |
Break (1069); High impedance (1291)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Review of the device ecog and stimulation path impedance in pdms confirmed that the ecog data was consistent with a lead break.
|
| |
|
Event or Problem Description
|
|
The patient had a depth lead removed and replaced on (b)(6) 2018.During a follow-up visit on (b)(6) 2018, it was noted that the left depth lead impedances were high.The treating clinician palpated the lead which produced artifact during real-time ecog recordings.A decision was made to replace the lead which demonstrated evidence of a possible lead break.The cause of the lead break is unknown.
|
| |
|