Blank fields on this form indicate the information is unknown, unchangaed, or unavailable.Additional information: upon preliminary investigation of the complaint device, an additional failure mode in the form of a split in the tip of the device was noted.Investigation ¿ evaluation.A review of the complaint history, device history record, drawing, instructions for use (ifu), manufactures instructions, quality control, specifications, and a visual inspection of the returned device were conducted during the investigation.The visual inspection of the returned package confirmed that the sheath and dilator of the rcfn-5.0-35-j set was returned for investigation.The tip of the dilator was flattened at 20cm from the hub for approximately 2.5mm in length.The endhole of the dilator appeared round and free of damage.No other damage was noted to the dilator.The tip of the sheath also appeared to have a small split in the tip.No other damage was noted to the sheath.Additionally, a document based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.A review of the device history record showed no nonconforming events which could contribute to this failure mode.However, a review of the component lot was reviewed and one potentially related nonconformances was discovered.This nonconformance was for surface damage, however the affected device was scrapped and not replaced.It should also be noted that the only complaint associated with this lot number is (b)(4) which was opened from this file to capture a second failure mode of the same device.Furthermore, a review of the manufactures instructions, quality control procedures, and overall device master record was conducted, and no gaps were discovered.Based on the information provided and the examination of the returned product, investigation has concluded that this event can likely be traced to the transportation and storage of the device.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment no further action is required.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|