| Model Number |
289706000 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
11/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that c5/6 acdf; dr.(b)(6), (b)(6) hospital, (b)(6) 2018.During final tightening of the cam lock on the uni-plate, the distal tip of the final tightener sheared off.The distal tip was then retrieved with a sucker and no adverse event occurred.The rep asked on two occasions if the distal tip had been retrieved and this was confirmed by the surgeon.Surgeon advised that it was sucked up and no longer in the patient.Action taken to manage the problem: jnj representative suggested to surgeon to use sucker to remove the distal tip of driver.Another driver was used to lock off/torque off.No ae to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was not returned for evaluation.A review of the device history record could not be performed as the lot number is unknown.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.As no systemic trends were found, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|