| Model Number |
EEAXL2835 |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Codes |
Failure to Anastomose (1028); No Code Available (3191)
|
| Date of Event |
11/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a laparoscopic colostomy closure, while creating the anastomosis, the stapler misfired.Another device was opened and the anastomosis had to be redone resulting in surgical extension of a few extra hours.Surgeon had to transect more of bowel.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information:evaluation summary post market vigilance (pmv) led an evaluation of one device.Visual inspection of the device under the microscope displayed nicks on the knife blade.Visual inspection of the cut ring noted that it was partially cut.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the observed knife blade damage may occur if the instrument is applied over an obstruction.The instructions for use manual for this product states: "4.Make certain that the section of tissue to be stapled is free from any metal or similar structure; otherwise, the knife blade may not cut."replication of the partial cut of the cut ring may occur if the instrument handle compression is not completed or if the instrument is applied against an excessive amount of tissue during the clinical application.Also, the instructions for use of this product states: when firing the stapler, make sure to fully squeeze the instrument handle until the metal underside of the handle contacts the stapler body to the fullest extent.Partial or incomplete handle squeezes may result in unacceptable staple formation and/or incomplete knife cut.The root cause of the observed damage was misuse of the product which caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|