• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SPINE INC BALL TIP FEELER STRAIGHT; PROBE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC BALL TIP FEELER STRAIGHT; PROBE Back to Search Results
Catalog Number 275010140
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that dr.(b)(6) was performing a posterior spinal fusion (scoliosis) and was using a ball tip probe from the expedium system.Upon palpating the pedicle with the ball tip probe, the ball tip probe broke in 2 pieces.The break occurred at the junction of the handle and the slimmer position of the shaft.The two pieces were retrieved.Nothing was left inside the patient.No delay was obtained.Given that the patient was not affected by this break, no patient information was obtained.I have nothing further to add.Patient consequence? : no.Is the information being submitted for this complaint all the details that are known/available regarding this event? : yes.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Udi: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was not returned for evaluation.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALL TIP FEELER STRAIGHT
Common Device Name
PROBE
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
MDR Report Key8161576
Report Number1526439-2018-51143
Device Sequence Number11355400
Product Code HXB
UDI-Device Identifier10705034194066
UDI-Public(01)10705034194066
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 11/26/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number275010140
Device Lot NumberTY17307
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 11/26/2018
Supplement Date Received by Manufacturer02/20/2019
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date02/20/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-