• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. RIO®SHELL IMPACT-OFFSET-TRIDENT PST; STEREOTAXIC DEVICE, ROBOTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. RIO®SHELL IMPACT-OFFSET-TRIDENT PST; STEREOTAXIC DEVICE, ROBOTICS Back to Search Results
Catalog Number 209830
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 11/26/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
¿as part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.¿.
 
Event or Problem Description
Offset impaction handle and impaction platform both broke upon impaction.Case type: tha.Surgical delay- =15 minutes.
 
Event or Problem Description
Offset impaction handle and impaction platform both broke upon impaction.Case type: tha, surgical delay- =15 minutes.
 
Additional Manufacturer Narrative
It was reported that the impactor broke during the impaction.The product was unavailable for inspection as the product was not returned.Product history review was not provided as the lot number was not provided.Complaint history review was not provided as the lot number was not provided.The failure could not be determined as the product was not available for inspection.No additional investigation or specific actions are required.No action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.Product was unavailable for inspection as the product was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RIO®SHELL IMPACT-OFFSET-TRIDENT PST
Common Device Name
STEREOTAXIC DEVICE, ROBOTICS
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
MDR Report Key8161731
Report Number3005985723-2018-00746
Device Sequence Number2126520
Product Code OLO
UDI-Device Identifier00848486030285
UDI-Public00848486030285
Combination Product (Y/N)N
Initial Reporter StateND
Initial Reporter CountryUS
PMA/510(K) Number
K141989
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number209830
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 11/26/2018
Supplement Date Received by Manufacturer01/29/2019
Initial Report FDA Received Date12/13/2018
Supplement Report FDA Received Date02/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
-
-